P Xiao, MEP de Jesus, Y Cui, RE Imhof & the TEWL Calibration Consortium.
Poster, World Congress on Non-Invasive Studies of the Skin, Philadelphia, September 2005.

 

Summary:-

The droplet method for TEWL calibration is a new technique, which has the advantages of gravimetric-based calibration, traceability and good repeatability. It can be used for calibrating any TEWL instruments that is capable of recording a time series of water vapour flux density readings. In this paper, we will present a mathematical analysis of calibration errors for both open-chamber and closed chamber TEWL instruments. The effects of sensor position, sensor calibration, chamber length, droplet position, and instrument response speed are analysed.

 

 